Current browse context:
physics.ins-det
Change to browse by:
References & Citations
Physics > Instrumentation and Detectors
Title: Development of a low-mass and high-efficiency charged particle detector
(Submitted on 14 Dec 2015 (v1), last revised 18 Dec 2015 (this version, v2))
Abstract: We developed a low-mass and high-efficiency charged particle detector for an experimental study of the rare decay $K_L \rightarrow \pi^0 \nu \bar{\nu}$. The detector is important to suppress the background with charged particles to the level below the signal branching ratio predicted by the Standard Model (O(10$^{-11}$)). The detector consists of two layers of 3-mm-thick plastic scintillators with wavelength shifting fibers embedded and Multi Pixel Photon Counters for readout. We manufactured the counter and evaluated the performance such as light yield, timing resolution, and efficiency. With this design, we achieved the inefficiency per layer against penetrating charged particles to be less than $1.5 \times 10^{-5}$, which satisfies the requirement of the KOTO experiment determined from simulation studies.
Submission history
From: Daichi Naito [view email][v1] Mon, 14 Dec 2015 03:51:28 GMT (1416kb)
[v2] Fri, 18 Dec 2015 01:22:39 GMT (1501kb)
Link back to: arXiv, form interface, contact.