Title: Neutron Dose measurement in the carbon ion therapy area at HIRFL (IMP) as 12C ions with energies from 165 to 350MeV/u

Abstract: The neutron dose distributions on observation distances and on observation angles were measured using a Wendi-II neutron dose-meter at the deep tumor therapy terminal at HIRFL (Heavy Ion Research Facility in Lanzhou) as 12C ions with energies from 165 to 350 MeV/u bombarding on thick solid water targets with different thickness according to the ion energies. The experimental results were compared with those calculated by FLUKA code. It is found that the experimental data was in good agreement with the calculated results. The neutron energy spectra were also studied by using the FLUKA code. The results are valuable for the shielding design of high energy heavy ion medical machines and for the individual dose assessment.
Subjects: Medical Physics (physics.med-ph); Instrumentation and Detectors (physics.ins-det)
Cite as: arXiv:1512.09246 [physics.med-ph]