We gratefully acknowledge support from
the Simons Foundation and member institutions.
Full-text links:

Download:

Current browse context:

physics

Change to browse by:

References & Citations

Bookmark

(what is this?)
CiteULike logo BibSonomy logo Mendeley logo del.icio.us logo Digg logo Reddit logo

Physics > Instrumentation and Detectors

Title: A 10 mK hermetic cell for eliminating parasitic heating in cryogen-free dilution refrigerators

Abstract: Cryogen-free dilution refrigerators generally simplify low temperature research but some types of samples, including superconducting qubits and other nanoelectronic devices, are affected by environmental heat sources such as stray photons or residual helium. We present the design and performance of a hermetic cell installed on the mixing chamber plate of a cryogen-free dilution refrigerator. The performance was quantified by measuring the dependence of the resonance frequency of a mechanical oscillator installed inside the cell on the mixing chamber temperature down to 10 mK. We found the expected logarithmic temperature dependence of the resonance frequency down to the lowest temperatures, demonstrating that the efficiency of the hermetic shield is significantly better than that of a simpler shield with no visible gaps.
Subjects: Instrumentation and Detectors (physics.ins-det); Applied Physics (physics.app-ph)
Cite as: arXiv:1911.03160 [physics.ins-det]
  (or arXiv:1911.03160v1 [physics.ins-det] for this version)

Submission history

From: Andrew Fefferman [view email]
[v1] Fri, 8 Nov 2019 10:08:20 GMT (354kb,D)

Link back to: arXiv, form interface, contact.