Current browse context:
physics.ins-det
Change to browse by:
References & Citations
Physics > Instrumentation and Detectors
Title: Performance and Quality Control of a Radiation-Hard 12-bit 40 MSPS ADC for the ATLAS Liquid Argon Calorimeter Trigger Readout Electronics Phase-I Upgrade at the LHC
(Submitted on 12 Dec 2019 (v1), revised 27 Feb 2020 (this version, v3), latest version 6 Mar 2020 (v4))
Abstract: A radiation-hard quad-channel 12-bit 40 MSPS pipeline analog-to-digital converter (ADC) has been designed for the trigger readout electronics Phase-I upgrade of the ATLAS Liquid Argon calorimeter, at the CERN Large Hadron Collider. The final version of the custom design, fabricated in a commercial 130 nm CMOS process, is presented and found to meet the system requirements for analog performance and radiation tolerance. The procedure for quality control of 17200 ADCs is described and the results are presented.
Submission history
From: Ines Ochoa [view email][v1] Thu, 12 Dec 2019 17:42:21 GMT (5049kb,D)
[v2] Mon, 16 Dec 2019 14:34:05 GMT (5049kb,D)
[v3] Thu, 27 Feb 2020 18:42:53 GMT (5050kb,D)
[v4] Fri, 6 Mar 2020 21:31:55 GMT (5050kb,D)
Link back to: arXiv, form interface, contact.