Current browse context:
physics.ins-det
Change to browse by:
References & Citations
Physics > Instrumentation and Detectors
Title: Value-assigned pulse shape discrimination for neutron detectors
(Submitted on 15 Jan 2020 (v1), last revised 18 Jun 2021 (this version, v2))
Abstract: Using the waveforms from a digital electronic system, an offline analysis technique on pulse shape discrimination (PSD) has been developed to improve the neutron-gamma separation in a bar-shaped NE-213 scintillator that couples to a photomultiplier tube (PMT) at each end. The new improved method, called the ``valued-assigned PSD'' (VPSD), assigns a normalized fitting residual to every waveform as the PSD value. This procedure then facilitates the incorporation of longitudinal position dependence of the scintillator, which further enhances the PSD capability of the detector system. In this paper, we use radiation emitted from an AmBe neutron source to demonstrate that the resulting neutron-gamma identification has been much improved when compared to the traditional technique that uses the geometric mean of light outputs from both PMTs. The new method has also been modified and applied to a recent experiment at the National Superconducting Cyclotron Laboratory (NSCL) that uses an analog electronic system.
Submission history
From: Fanurs Chi En Teh [view email][v1] Wed, 15 Jan 2020 21:17:13 GMT (2679kb,D)
[v2] Fri, 18 Jun 2021 01:03:24 GMT (5428kb,D)
Link back to: arXiv, form interface, contact.