Title: Thermal Kinetic Inductance Detectors for millimeter-wave detection
(Submitted on 24 Jan 2020 (v1), last revised 29 Jul 2020 (this version, v2))
Abstract: Thermal Kinetic Inductance Detectors (TKIDs) combine the excellent noise performance of traditional bolometers with a radio frequency multiplexing architecture that enables the large detector counts needed for the next generation of millimeter-wave instruments. In this paper, we first discuss the expected noise sources in TKIDs and derive the limits where the phonon noise contribution dominates over the other detector noise terms: generation-recombination, amplifier, and two-level system (TLS) noise. Second, we characterize aluminum TKIDs in a dark environment. We present measurements of TKID resonators with quality factors of about $10^5$ at 80 mK. We also discuss the bolometer thermal conductance, heat capacity, and time constants. These were measured by the use of a resistor on the thermal island to excite the bolometers. These dark aluminum TKIDs demonstrate a noise equivalent power NEP = $2 \times 10^{-17} \mathrm{W}/\mathrm{\sqrt{Hz}} $, with a $1/f$ knee at 0.1 Hz, which provides background noise limited performance for ground-based telescopes observing at 150 GHz.
Submission history
From: Albert Wandui [view email][v1] Fri, 24 Jan 2020 05:27:11 GMT (7178kb)
[v2] Wed, 29 Jul 2020 21:41:33 GMT (6938kb)