We gratefully acknowledge support from
the Simons Foundation and member institutions.
Full-text links:

Download:

Current browse context:

physics.ins-det

Change to browse by:

References & Citations

Bookmark

(what is this?)
CiteULike logo BibSonomy logo Mendeley logo del.icio.us logo Digg logo Reddit logo

Physics > Instrumentation and Detectors

Title: Automated laser induced damage threshold testing applied to a ferroelectric spatial light modulator

Abstract: Laser Induced Damage Thresholds (LIDTs) are a measure of the level of fluence an optical component may be expected to handle without observable damage. In this work we present an automated system for measurement of LIDTs for a wide range of components and include details of the data handling, image processing and code required. We then apply this system to LIDT measurements of a commercial HDP-1280-2 'BlueJay' ferroelectric display finding a LIDT of 9.2 W/cm2 {\O}27{\mu}m, 5.5 W/cm2 {\O}150{\mu}m and 3.2 W/cm2 {\O}3.1mm with wavelength 1090+-5nm. Finally, the quality of the results obtained are discussed and conclusions drawn.
Subjects: Instrumentation and Detectors (physics.ins-det)
Cite as: arXiv:2004.00712 [physics.ins-det]
  (or arXiv:2004.00712v1 [physics.ins-det] for this version)

Submission history

From: Peter Christopher [view email]
[v1] Mon, 30 Mar 2020 10:17:32 GMT (5794kb,D)

Link back to: arXiv, form interface, contact.