We gratefully acknowledge support from
the Simons Foundation and member institutions.
Full-text links:

Download:

Current browse context:

physics.ins-det

Change to browse by:

References & Citations

Bookmark

(what is this?)
CiteULike logo BibSonomy logo Mendeley logo del.icio.us logo Digg logo Reddit logo

Physics > Instrumentation and Detectors

Title: Laser Calibration System for Time of Flight Scintillator Arrays

Abstract: A laser calibration system was developed for monitoring and calibrating time of flight (TOF) scintillating detector arrays. The system includes setups for both small- and large-scale scintillator arrays. Following test-bench characterization, the laser system was recently commissioned in experimental Hall B at the Thomas Jefferson National Accelerator Facility for use on the new Backward Angle Neutron Detector (BAND) scintillator array. The system successfully provided time walk corrections, absolute time calibration, and TOF drift correction for the scintillators in BAND. This showcases the general applicability of the system for use on high-precision TOF detectors.
Comments: 11 pages, 11 figures
Subjects: Instrumentation and Detectors (physics.ins-det); Nuclear Experiment (nucl-ex)
DOI: 10.1016/j.nima.2020.164177
Cite as: arXiv:2004.10268 [physics.ins-det]
  (or arXiv:2004.10268v2 [physics.ins-det] for this version)

Submission history

From: Axel Schmidt [view email]
[v1] Tue, 21 Apr 2020 20:02:10 GMT (4111kb,D)
[v2] Thu, 21 May 2020 14:58:10 GMT (5097kb,D)

Link back to: arXiv, form interface, contact.