Title: Using $^{22}$Na and $^{83{\rm m}}$Kr to calibrate and study the properties of scintillation in xenon-doped liquid argon
(Submitted on 3 Jul 2020 (v1), last revised 26 Oct 2020 (this version, v3))
Abstract: We have measured the properties of scintillation light in liquid argon doped with xenon concentrations from 165 ppm to 10,010 ppm using a $^{22}$Na source. The energy transfer processes in the xenon-doped liquid argon are discussed in detail, and a new waveform model is established and used to fit the average waveform. The time profile of the scintillation photon in the xenon-doped liquid argon and of the TPB emission are presented. The quantities of xenon-doped are controlled by a Mass Flow Controller which is calibrated via a Redusial Gas Analyzer to ensure that the xenon concentration is accurate. In addition, a successful test of $^{83{\rm m}}$Kr as a calibration source has been implemented in the xenon-doped liquid argon detector for the first time. By comparing the light yield of the $^{22}$Na and $^{83{\rm m}}$Kr, it can be concluded that the scintillation efficiency is almost same over the range of 41.5 keV to 511 keV.
Submission history
From: Yongpeng Zhang [view email][v1] Fri, 3 Jul 2020 08:49:14 GMT (362kb,D)
[v2] Mon, 20 Jul 2020 13:48:39 GMT (581kb,D)
[v3] Mon, 26 Oct 2020 02:00:14 GMT (827kb,D)
Link back to: arXiv, form interface, contact.