Current browse context:
physics.ins-det
Change to browse by:
References & Citations
Physics > Instrumentation and Detectors
Title: Performance of a Large Area Photon Detector For Rare Event Search Applications
(Submitted on 29 Sep 2020 (this version), latest version 11 Jan 2021 (v2))
Abstract: We present the design and characterization of a large-area Cryogenic PhotoDetector (CPD) designed for active particle identification in rare event searches, such as neutrinoless double beta decay and dark matter experiments. The detector consists of a $45.6$ $\mathrm{cm}^2$ surface area by 1-mm-thick $10.6$ $\mathrm{g}$ Si wafer. It is instrumented with a distributed network of Quasiparticle-trap-assisted Electrothermal feedback Transition-edge sensors (QETs) with $T_c=41.5$ $\mathrm{mK}$ to measure athermal phonons released from interactions with photons. The detector is characterized and calibrated in the center of the detector with a collimated $^{55}$Fe X-ray source. The noise equivalent power is measured to be $1\times 10^{-17}$ $\mathrm{W}/\sqrt{\mathrm{Hz}}$ in a bandwidth of $2.7$ $\mathrm{kHz}$. The baseline energy resolution is measured to be $\sigma_E = 3.86 \pm 0.04$ $(\mathrm{stat.})^{+0.23}_{-0.00}$ $(\mathrm{syst.})$ $\mathrm{eV}$ (RMS). The detector also has an expected timing resolution of $\sigma_t = 2.3$ $\mu\mathrm{s}$ for $5$ $\sigma_E$ events.
Submission history
From: Samuel Watkins [view email][v1] Tue, 29 Sep 2020 20:49:57 GMT (622kb,D)
[v2] Mon, 11 Jan 2021 16:40:55 GMT (626kb,D)
Link back to: arXiv, form interface, contact.