Title: Development of a $^{83\mathrm{m}}$Kr source for the calibration of the CENNS-10 Liquid Argon Detector

Authors: COHERENT Collaboration: D. Akimov, P. An, C. Awe, P. S. Barbeau, B. Becker, V. Belov, I. Bernardi, M. A. Blackston, L. Blokland, A. Bolozdynya, B. Cabrera-Palmer, N. Chen, D. Chernyak, E. Conley, J. Daughhetee, M. del Valle Coello, J. A. Detwiler, M. R. Durand, Y. Efremenko, S. R. Elliott, L. Fabris, M. Febbraro, W. Fox, A. Galindo-Uribarri, A. Gallo Rosso, M. P. Green, K. S. Hansen, M. R. Heath, S. Hedges, M. Hughes, T. Johnson, A. Khromov, A. Konovalov, E. Kozlova, A. Kumpan, L. Li, J. T. Librande, J. M. Link, J. Liu, K. Mann, D. M. Markoff, O. McGoldrick, P. E. Mueller, J. Newby, D. S. Parno, S. Pentilla, D. Pershey, D. Radford, R. Rapp, H. Ray, J. Raybern, O. Razuvaeva, D. Reyna, G. C. Rich, D. Rudik, J. Runge, D. J. Salvat, K. Scholberg, A. Shakirov, G. Simakov, W. M. Snow, V. Sosnovtsev, et al. (17 additional authors not shown)
Abstract: We report on the preparation of and calibration measurements with a $^{83\mathrm{m}}$Kr source for the CENNS-10 liquid argon detector. $^{83\mathrm{m}}$Kr atoms generated in the decay of a $^{83}$Rb source were introduced into the detector via injection into the Ar circulation loop. Scintillation light arising from the 9.4 keV and 32.1 keV conversion electrons in the decay of $^{83\mathrm{m}}$Kr in the detector volume were then observed. This calibration source allows the characterization of the low-energy response of the CENNS-10 detector and is applicable to other low-energy-threshold detectors. The energy resolution of the detector was measured to be 9$\%$ at the total $^{83\mathrm{m}}$Kr decay energy of 41.5 keV. We performed an analysis to separately calibrate the detector using the two conversion electrons at 9.4 keV and 32.1 keV
Comments: v2: As accepted to JINST
Subjects: Instrumentation and Detectors (physics.ins-det); High Energy Physics - Experiment (hep-ex); Nuclear Experiment (nucl-ex)
Journal reference: JINST 16 P04002 (2021)
DOI: 10.1088/1748-0221/16/04/P04002
Cite as: arXiv:2010.11258 [physics.ins-det]
  (or arXiv:2010.11258v2 [physics.ins-det] for this version)

Submission history

From: Jacob Zettlemoyer [view email]
[v1] Wed, 21 Oct 2020 19:27:16 GMT (3248kb,D)
[v2] Thu, 28 Jan 2021 00:03:19 GMT (3704kb,D)

Link back to: arXiv, form interface, contact.