References & Citations
Quantitative Biology > Quantitative Methods
Title: Beat-to-Beat Fetal Heart Rate Analysis Using Portable Medical Device and Wavelet Transformation Technique
(Submitted on 1 Mar 2021 (v1), last revised 27 Jun 2022 (this version, v4))
Abstract: A beat-to-beat Tele-fetal Monitoring and comparison with clinical data are studied with a wavelet transformation approach. Tele-fetal monitoring is a big progress toward a wearable medical device for a pregnant woman capable of obtaining prenatal care at home. We apply a wavelet transformation algorithm for fetal cardiac monitoring using a portable fetal Doppler medical device. Choosing an appropriate mother wavelet, 85 different mother wavelets are investigated. The efficiency of the proposed method is evaluated using two data sets including public and clinical. From publicly available data on PhysioBank, and simultaneous clinical measurement, we prove that the comparison between obtained fetal heart rate by the algorithm and the baselines yields a promising accuracy beyond 95%. Finally, we conclude that the proposed algorithm would be a robust technique for any similar tele-fetal monitoring approach.
Submission history
From: Maria Farahi [view email][v1] Mon, 1 Mar 2021 14:02:24 GMT (1786kb,D)
[v2] Wed, 3 Mar 2021 10:43:19 GMT (1786kb,D)
[v3] Thu, 4 Nov 2021 16:38:31 GMT (7729kb,D)
[v4] Mon, 27 Jun 2022 11:49:49 GMT (2079kb,D)