Title: Stationary Multi-source AI-powered Real-time Tomography (SMART) for Dynamic Cardiac Imaging

Abstract: A first stationary multi-source computed tomography (CT) system is prototyped for preclinical imaging to achieve real-time temporal resolution for dynamic cardiac imaging. This unique is featured by 29 source-detector pairs fixed on a circular track for each detector to collect x-ray signals only from the opposite x-ray source. The new system architecture potentially leads to a major improvement in temporal resolution. To demonstrate the feasibility of this Stationary Multi-source AI-based Real-time Tomography (SMART) system, we develop a novel reconstruction scheme integrating both sparsified image prior (SIP) and deep image prior (DIP), which is referred to as the SIP-DIP network. Then, the SIP-DIP network for cardiac imaging is evaluated on preclinical cardiac datasets of alive rats. The reconstructed image volumes demonstrate the feasibility of the SMART system and the SIP-DIP network and the merits over other reconstruction methods.
Comments: 15 pages, 6 figures, 30 references
Subjects: Quantitative Methods (q-bio.QM); Image and Video Processing (eess.IV)
Cite as: arXiv:2108.12076 [q-bio.QM]
  (or arXiv:2108.12076v1 [q-bio.QM] for this version)

Submission history

From: Ge Wang Dr. [view email]
[v1] Fri, 27 Aug 2021 01:00:47 GMT (1320kb)
[v2] Tue, 8 Feb 2022 03:33:39 GMT (3051kb)

Link back to: arXiv, form interface, contact.