Title: Quality assurance test and Failure Analysis of SiPM Arrays of GECAM Satellites

Abstract: The Gravitational wave high-energy Electromagnetic Counterpart All-sky Monitor (GECAM) satellite consists of two small satellites. Each GECAM payload contains 25 gamma ray detectors (GRD) and 8 charged particle detectors (CPD). GRD is the main detector which can detect gamma-rays and particles and localize the Gamma-Ray Bursts (GRB),while CPD is used to help GRD to discriminate gamma-ray bursts and charged particle bursts. The GRD makes use of lanthanum bromide (LaBr3) crystal readout by SiPM. As the all available SiPM devices belong to commercial grade, quality assurance tests need to be performed in accordance with the aerospace specifications. In this paper, we present the results of quality assurance tests, especially a detailed mechanism analysis of failed devices during the development of GECAM. This paper also summarizes the application experience of commercial-grade SiPM devices in aerospace payloads, and provides suggestions for forthcoming SiPM space applications.
Comments: 13 pages, 23 figures
Subjects: Instrumentation and Detectors (physics.ins-det)
Journal reference: RDTM-D-21-00057R4.2021.9.1
Cite as: arXiv:2109.00235 [physics.ins-det]
  (or arXiv:2109.00235v2 [physics.ins-det] for this version)

Submission history

From: Da-Li Zhang [view email]
[v1] Wed, 1 Sep 2021 08:14:51 GMT (1715kb)
[v2] Thu, 9 Dec 2021 09:00:33 GMT (1716kb)