Current browse context:
physics.med-ph
Change to browse by:
References & Citations
Physics > Medical Physics
Title: A Bayesian Optimization Approach for Attenuation Correction in SPECT Brain Imaging
(Submitted on 24 Sep 2021)
Abstract: Photon attenuation and scatter are the two main physical factors affecting the diagnostic quality of SPECT in its applications in brain imaging. In this work, we present a novel Bayesian Optimization approach for Attenuation Correction (BOAC) in SPECT brain imaging. BOAC utilizes a prior model parametrizing the head geometry and exploits High Performance Computing (HPC) to reconstruct attenuation corrected images without requiring prior anatomical information from complementary CT scans. BOAC is demonstrated in SPECT brain imaging using noisy and attenuated sinograms, simulated from numerical phantoms. The quality of the tomographic images obtained with the proposed method are compared to those obtained without attenuation correction by employing the appropriate image quality metrics. The quantitative results show the capacity of BOAC to provide images exhibiting higher contrast and less background artifacts as compared to the non-attenuation corrected MLEM images.
Submission history
From: Loizos Koutsantonis [view email][v1] Fri, 24 Sep 2021 12:27:06 GMT (637kb,D)
Link back to: arXiv, form interface, contact.