Title: Frequency-Dependent F-Number Increases the Contrast and the Spatial Resolution in Fast Pulse-Echo Ultrasound Imaging
(Submitted on 8 Nov 2021 (v1), last revised 5 Jan 2023 (this version, v2))
Abstract: Fixed $F$-numbers reduce grating lobe artifacts in fast pulse-echo ultrasound imaging. Such $F$-numbers result in dynamic receive subapertures whose widths vary with the focal position. These subapertures, however, ignore useful low-frequency components in the excluded radio frequency (RF) signals and, thus, reduce the lateral resolution. Here, we propose a frequency-dependent $F$-number to simultaneously suppress grating lobe artifacts and maintain the lateral resolution. This $F$-number, at high frequencies, reduces the receive subaperture to remove spatially undersampled components of the RF signals and suppress grating lobes. The $F$-number, at low frequencies, enlarges the receive subaperture to use the components of all RF signals and maintain the lateral resolution. Experiments validated the proposed $F$-number and demonstrated improvements in the contrast and the widths of wire targets of up to 3.2 % and 12.8 %, respectively.
Submission history
From: Martin Schiffner [view email][v1] Mon, 8 Nov 2021 15:57:49 GMT (2171kb,D)
[v2] Thu, 5 Jan 2023 13:21:00 GMT (2171kb,D)