We gratefully acknowledge support from
the Simons Foundation and member institutions.
Full-text links:

Download:

Current browse context:

physics.ins-det

Change to browse by:

References & Citations

Bookmark

(what is this?)
CiteULike logo BibSonomy logo Mendeley logo del.icio.us logo Digg logo Reddit logo

Physics > Instrumentation and Detectors

Title: Fluke 8588A and Keysight 3458A DMM Sampling Performance

Authors: Rado Lapuh (1), Jan Kucera (2), Jakub Kovac (2), Bostjan Voljc (3) ((1) Metrology Institute of the Republic of Slovenia, Celje, Slovenia, (2) Czech Metrology Institute, Brno, Czech Republic, (3) Slovenian Institute of Quality and Metrology, Ljubljana, Slovenia)
Abstract: This paper reports on an evaluation of the Fluke 8588A digital multimeter (DMM) sampling performance and comparison to Keysight 3458A DMM sampling performance. The design of 8588A type DMM shows both similarities and striking differences in design compared to 3458A type DMM. Apart from their specified sampling capabilities, measurements were performed to evaluate frequency response, noise performance, including a 1 / f type noise and a timing jitter induced noise, as well as phase measurement performance. The given comprehensive report is crucial for careful application of DMMs in metrology tasks with optimal utilization of their performance.
Comments: 11 pages, 26 figures
Subjects: Instrumentation and Detectors (physics.ins-det)
Cite as: arXiv:2205.11321 [physics.ins-det]
  (or arXiv:2205.11321v1 [physics.ins-det] for this version)

Submission history

From: Jan Kučera [view email]
[v1] Mon, 23 May 2022 14:06:53 GMT (2175kb,D)

Link back to: arXiv, form interface, contact.