Current browse context:
physics.ins-det
Change to browse by:
References & Citations
Physics > Instrumentation and Detectors
Title: Study on SiPM performance at low temperatures between $-60^{\circ}$C and $-20^{\circ}$C
(Submitted on 13 Jul 2022 (v1), last revised 26 Oct 2022 (this version, v4))
Abstract: Radon is the main background source of dark matter and neutrino experiments. Radon concentration ($\rm mBq/m^3$) measurement by liquid scintillation detector is a highly sensitive method at low temperatures using silicon photomultipliers (SiPMs) arrays. The SiPM performance characteristics are closely related to the lower detection limit of the detector. In this study, we built an automatic and accurate low-temperature measurement system to study the single photoelectron spectrum, SPE resolution, optical crosstalk, and after-pulse of the SiPM at different temperatures. As a result, we obtained the variation trend of the SiPM parameters at different temperatures, and the SiPM optimal working conditions were obtained, which can improve the detector's sensitivity
Submission history
From: Cui Zhong [view email][v1] Wed, 13 Jul 2022 12:23:26 GMT (5340kb)
[v2] Fri, 19 Aug 2022 14:18:01 GMT (5383kb)
[v3] Mon, 3 Oct 2022 12:55:26 GMT (5399kb)
[v4] Wed, 26 Oct 2022 09:02:38 GMT (5399kb)
Link back to: arXiv, form interface, contact.