Title: Improvement of primary power standard through international comparison feedback

Abstract: The high frequency primary power standard is univocally realized by means of a coaxial microcalorimeter, at least up to 40 GHz. The coaxial microcalorimeter is a broadband measurement system adjusted for effec-tive efficiency measurement of power sensors both of bolometric and thermoelectric type. The critical point in the power standard realization is in the determination of the calibration constant g of the microcalorimeter, a frequency dependent parameter that has huge impact on the accuracy as-sessment of the standard. The paper proposes a simple but powerful way for improving this accuracy by using the reference values provided by in-ternational key-comparisons which the microcalorimeter was involved in.
Comments: 12 pages, 1 figure, 2 tables
Subjects: Instrumentation and Detectors (physics.ins-det); Applied Physics (physics.app-ph)
Journal reference: LUCIANO BRUNETTI, LUCA OBERTO, MARCO SELLONE, Measurement 42 (2009) 1487-1490
DOI: 10.1016/j.measurement.2009.07.009