Title: Open-bore vertical MRI scanners generate significantly less RF heating around deep brain stimulation leads compared to horizontal scanners

Authors: Jasmine Vu (1 and 2), Bhumi Bhusal (2), Bach T Nguyen (2), Pia Sanpitak (1 and 2), Elizabeth Nowac (3), Julie Pilitsis (4), Joshua Rosenow (5), Laleh Golestanirad (1,2) ((1) Department of Biomedical Engineering, McCormick School of Engineering, Northwestern University, (2) Department of Radiology, Feinberg School of Medicine, Northwestern University, (3) Illinois Bone and Joint Institute, (4) Department of Neurosciences & Experimental Therapeutics, Albany Medical College, and (5) Department of Neurosurgery, Feinberg School of Medicine, Northwestern University)
Abstract: Objectives Studies that assess magnetic resonance imaging (MRI) induced radiofrequency (RF) heating of the tissue in the presence of an active electronic implant are mostly performed in horizontal, closed-bore scanners. Vertical, open-bore MRI systems have a 90{\deg} rotated magnet and generate a fundamentally different RF field distribution in the body, yet little is known about the RF heating of deep brain stimulation (DBS) systems in this class of scanners. Here, we investigated whether RF heating of DBS devices was significantly different in a vertical, open-bore MRI scanner compared to a horizontal, closed-bore MRI scanner. Materials and Methods In this phantom study, RF heating around the lead of a commercial DBS system implanted in an anthropomorphic phantom was evaluated in a 1.2 T vertical open-bore scanner (Oasis, Fujifilm Healthcare) and a 1.5 T horizontal closed-bore scanner (Aera, Siemens Healthineers). DBS devices were implanted following 30 realistic lead trajectories. Electromagnetic simulations were performed to assess the specific absorption rate (SAR) of RF energy around leads with different internal structures. Results When controlling for B1+rms, temperature increase around the lead-tip was significantly lower at the vertical scanner compared to the horizontal scanner (p-value=9.1x10-7). Electromagnetic simulations demonstrated up to a 14-fold reduction in the maximum 0.1g-averaged SAR deposited in the tissue surrounding the lead-tip in a vertical scanner compared to a horizontal scanner for leads with straight and helical internal wires. Conclusions RF experiments and electromagnetic simulations demonstrated consistently lower RF heating and power deposition around the DBS lead-tip at the vertical scanner compared to the horizontal scanner. Our simulation results suggest that this trend in heating may potentially extend to leads from other manufacturers.
Comments: 29 pages, 10 figures