Title: The Highly Miniaturised Radiation Monitor
(Submitted on 15 Jan 2014 (v1), last revised 20 Jun 2014 (this version, v2))
Abstract: We present the design and preliminary calibration results of a novel highly miniaturised particle radiation monitor (HMRM) for spacecraft use. The HMRM device comprises a telescopic configuration of active pixel sensors enclosed in a titanium shield, with an estimated total mass of 52 g and volume of 15 cm$^3$. The monitor is intended to provide real-time dosimetry and identification of energetic charged particles in fluxes of up to 10$^8$ cm$^{-2}$ s$^{-1}$ (omnidirectional). Achieving this capability with such a small instrument could open new prospects for radiation detection in space.
Submission history
From: Edward Mitchell [view email][v1] Wed, 15 Jan 2014 09:57:58 GMT (1004kb,D)
[v2] Fri, 20 Jun 2014 17:35:32 GMT (1005kb,D)