Physics > Instrumentation and Detectors

Title: The Highly Miniaturised Radiation Monitor

Abstract: We present the design and preliminary calibration results of a novel highly miniaturised particle radiation monitor (HMRM) for spacecraft use. The HMRM device comprises a telescopic configuration of active pixel sensors enclosed in a titanium shield, with an estimated total mass of 52 g and volume of 15 cm$^3$. The monitor is intended to provide real-time dosimetry and identification of energetic charged particles in fluxes of up to 10$^8$ cm$^{-2}$ s$^{-1}$ (omnidirectional). Achieving this capability with such a small instrument could open new prospects for radiation detection in space.
Comments: 17 pages, 15 figures
Subjects: Instrumentation and Detectors (physics.ins-det)
DOI: 10.1088/1748-0221/9/07/P07010
Cite as: arXiv:1401.3533 [physics.ins-det]