Physics > Instrumentation and Detectors

Title: An Energy-dependent Electro-thermal Response Model of CUORE Cryogenic Calorimeter

Authors: CUORE Collaboration: D. Q. Adams, C. Alduino, K. Alfonso, F. T. Avignone III, O. Azzolini, G. Bari, F. Bellini, G. Benato, M. Beretta, M. Biassoni, A. Branca, C. Brofferio, C. Bucci, J. Camilleri, A. Caminata, A. Campani, L. Canonica, X. G. Cao, S. Capelli, C. Capelli, L. Cappelli, L. Cardani, P. Carniti, N. Casali, E. Celi, D. Chiesa, M. Clemenza, S. Copello, O. Cremonesi, R. J. Creswick, A. D'Addabbo, I. Dafinei, F. Del Corso, S. Dell'Oro, S. Di Domizio, S. Di Lorenzo, V. Dompè, D. Q. Fang, G. Fantini, M. Faverzani, E. Ferri, F. Ferroni, E. Fiorini, M. A. Franceschi, S. J. Freedman, S.H. Fu, B. K. Fujikawa, S. Ghislandi, A. Giachero, A. Gianvecchio, L. Gironi, A. Giuliani, P. Gorla, C. Gotti, T. D. Gutierrez, K. Han, E. V. Hansen, K. M. Heeger, R. G. Huang, H. Z. Huang, J. Johnston, et al. (59 additional authors not shown)
Abstract: The Cryogenic Underground Observatory for Rare Events (CUORE) is the most sensitive experiment searching for neutrinoless double-beta decay ($0\nu\beta\beta$) in $^{130}\text{Te}$. CUORE uses a cryogenic array of 988 TeO$_2$ calorimeters operated at $\sim$10 mK with a total mass of 741 kg. To further increase the sensitivity, the detector response must be well understood. Here, we present a non-linear thermal model for the CUORE experiment on a detector-by-detector basis. We have examined both equilibrium and dynamic electro-thermal models of detectors by numerically fitting non-linear differential equations to the detector data of a subset of CUORE channels which are well characterized and representative of all channels. We demonstrate that the hot-electron effect and electric-field dependence of resistance in NTD-Ge thermistors alone are inadequate to describe our detectors' energy dependent pulse shapes. We introduce an empirical second-order correction factor in the exponential temperature dependence of the thermistor, which produces excellent agreement with energy-dependent pulse shape data up to 6 MeV. We also present a noise analysis using the fitted thermal parameters and show that the intrinsic thermal noise is negligible compared to the observed noise for our detectors.
Comments: 34 pages, 14 figures, 6 tables
Subjects: Instrumentation and Detectors (physics.ins-det); Nuclear Experiment (nucl-ex)
Cite as: arXiv:2205.04549 [physics.ins-det]
  (or arXiv:2205.04549v2 [physics.ins-det] for this version)

Submission history

From: Mingyu Li [view email]
[v1] Mon, 9 May 2022 20:51:51 GMT (1401kb,D)
[v2] Thu, 28 Jul 2022 20:28:59 GMT (1406kb,D)

Link back to: arXiv, form interface, contact.