Title: A Cryogenic Torsion Balance Using a Liquid-Cryogen Free, Ultra-Low Vibration Cryostat

Authors: S. M. Fleischer (1), M. P. Ross (2), K. Venkateswara (2), C. A. Hagedorn (2), E. A. Shaw (2), E. Swanson (2), B. R.Heckel (2), J. H. Gundlach (2) ((1) Western Washington University, (2) University of Washington)
Abstract: We describe a liquid-cryogen free cryostat with ultra-low vibration levels which allows for continuous operation of a torsion balance at cryogenic temperatures. The apparatus uses a commercially available two-stage pulse-tube cooler and passive vibration isolation. The torsion balance exhibits torque noise levels lower than room temperature thermal noise by a factor of about four in the frequency range of 3-10mHz, limited by residual seismic motion and by radiative heating of the pendulum body. In addition to lowering thermal noise below room-temperature limits, the low-temperature environment enables novel torsion balance experiments. Currently, the maximum duration of a continuous measurement run is limited by accumulation of cryogenic surface contamination on the optical elements inside the cryostat.
Comments: 8 pages, 6 figures
Subjects: Instrumentation and Detectors (physics.ins-det)
Journal reference: Review of Scientific Instruments 93, 064505 (2022)
DOI: 10.1063/5.0089933
Cite as: arXiv:2206.02890 [physics.ins-det]