Physics > Instrumentation and Detectors

Title: Response of a Liquid $^3$He Neutron Detector

Abstract: The $^3$He(n,p) process is excellent for neutron detection between thermal and $\sim$4\,MeV because of the high cross section and near-complete energy transfer from the neutron to the proton. This process is typically used in gaseous forms with ionization readout detectors. Here we study the response of a liquid $^3$He neutron detector with a scintillation readout. We anticipate an efficiency boost of around a factor of 64 compared to 10\,atm gaseous detectors, given similar detector volumes.
Comments: 10 pages, 1 table, 8 figures
Subjects: Instrumentation and Detectors (physics.ins-det)
DOI: 10.1063/10.0022475