Title: Optimization of Dark-Field CT for Lung Imaging

Abstract: Background: X-ray grating-based dark-field imaging can sense the small angle scattering caused by an object's micro-structure. This technique is sensitive to lung's porous alveoli and is able to detect lung disease at an early stage. Up to now, a human-scale dark-field CT has been built for lung imaging. Purpose: This study aimed to develop a more thorough optimization method for dark-field lung CT and summarize principles for system design. Methods: We proposed a metric in the form of contrast-to-noise ratio (CNR) for system parameter optimization, and designed a phantom with concentric circle shape to fit the task of lung disease detection. Finally, we developed the calculation method of the CNR metric, and analyzed the relation between CNR and system parameters. Results: We showed that with other parameters held constant, the CNR first increases and then decreases with the system auto-correlation length (ACL). The optimal ACL is nearly not influenced by system's visibility, and is only related to phantom's property, i.e., scattering material's size and phantom's absorption. For our phantom, the optimal ACL is about 0.21 {\mu}m. As for system geometry, larger source-detector and isocenter-detector distance can increase the system's maximal ACL, helping the system meet the optimal ACL more easily. Conclusions: This study proposed a more reasonable metric and a task-based process for optimization, and demonstrated that the system optimal ACL is only related to the phantom's property.
Comments: There is a mistake in subsection 2.3, where the content is not correct because of the incorrect parameter we set, which leads to the following calculations in the following sections potentially incorrect
Subjects: Medical Physics (physics.med-ph); Image and Video Processing (eess.IV)
Cite as: arXiv:2405.00259 [physics.med-ph]
  (or arXiv:2405.00259v2 [physics.med-ph] for this version)

Submission history

From: Peiyuan Guo [view email]
[v1] Wed, 1 May 2024 00:24:23 GMT (476kb)
[v2] Thu, 2 May 2024 02:32:32 GMT (0kb,I)

Link back to: arXiv, form interface, contact.